PM-60A Touch Screen Hand Held Oximeter

The PM60A Handheld Pulse Oximeter is a hospital quality patient monitor.   This unit is a touch screen pulse oximeter that has a Routine Check Mode and Continuous Monitoring Mode, Data Graph and Trend Table Review, Rich Analysis Report, and more!

More details


$225.00

PM-60A

Availability: This product is no longer in stock with these options but is available with others


Notify me when available

1 other products in the same category:

Features of the PM60A Handheld Pulse Oximeter:
- Routine Check Mode and Continuous Monitoring Mode
- Data Graph and Trend Table Review
- Rich Analysis Report
- Perfect Mount Solution
- Convenient operation by Touch Screen
- Large SD card memory for more than 10,000 patient case storage
- Advanced analysis software for sleep study (Free with the oximeter)
- Sync with PC based Software.  Not compatible with Mac!!!
- Note that finger probe contains Latex!!

Specifications for the PM-60A Handheld Pulse Oximeter:
- Display: 3.5" Color TFT
- Resolution: 320x240
- Display Mode: Standard face, Waveform face, Display Direction adjustable  
- Indicator: Power indicator light, Alarm sound, Pulse tone
- Interface: One dual-purpose socket for connecting SPO2 sensors and
    communication cables
- Power Supply: DC 5V
- Battery: Built in Lithium Ion, 2 hours for charging, 8 hours for continuous
    working.
- Trend Graph: Resolution from 1s, 5s, 10s. Maximum time 96 hours.
- Trend Table: Resolution from 1s, 5s, 10s.  Review up to 2000 items.
- Alarm: Adjustable High and Low limits. Three level audible and visual
    alarms.

Technical specifications for the PM-60A Handheld Pulse Oximeter:
- Meets the requirement of IEC60601 series
- Type of Protection: Class II with internal electric power supply  
- Dimension: 92mm(W) x 82mm(H) x 22mm(D)
- Weight: 136g (with battery)
- Temperature: 0?~ +40?
- Humidity: 15% ~ 95%
- Temperature: -20?~ +60?
- Humidity: 10% ~ 95%
- Neonate, Pediatric and  Adult

SPO2
- Measurement Range: 0 ~ 100%
- Resolution: 1%
- Accuracy:   ±2%(70%~100%, Adult/Pediatric)
                       ±3%(70%~100%, Neonate)
                       0% ~ 69% unspecified
- Alarm Range: 0% ~ 100%
- Refreshing Rate: 1 second

Pulse
- Pulse Rate: Measurement Range: 25 ~ 250 bpm
- Resolution: 1bpm
- Accuracy: ±1bpm
- Alarm Range: 25 ~ 250 bmp
- Refreshing Rate: 1s

Valid XHTML 1.0 Transitional Valid CSS!